| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.105 | 2.038 |
| 50.00 | 1.627 | 1.560 |
| 25.00 | 1.239 | 1.172 |
| 12.50 | 0.867 | 0.800 |
| 6.25 | 0.576 | 0.509 |
| 3.13 | 0.321 | 0.254 |
| 1.56 | 0.240 | 0.173 |
| 0.00 | 0.067 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-118% | 106% |
| EDTA plasma(n=5) | 80-89% | 85% |
| Heparin plasma(n=5) | 88-109% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-97% | 90-112% | 94-102% | 87-99% |
| EDTA plasma(n=5) | 92-109% | 89-108% | 89-101% | 87-110% |
| Heparin plasma(n=5) | 87-113% | 82-107% | 84-95% | 80-96% |