| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 5000.00 | 2.157 | 2.089 | 
| 2500.00 | 1.563 | 1.495 | 
| 1250.00 | 1.096 | 1.028 | 
| 625.00 | 0.887 | 0.819 | 
| 312.50 | 0.582 | 0.514 | 
| 156.25 | 0.329 | 0.261 | 
| 78.13 | 0.179 | 0.111 | 
| 0.00 | 0.068 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 80-93% | 86% | 
| EDTA plasma(n=5) | 87-97% | 90% | 
| Heparin plasma(n=5) | 88-102% | 95% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-99% | 98-105% | 96-101% | 
| EDTA plasma(n=5) | 95-103% | 91-105% | 78-90% | 89-97% | 
| Heparin plasma(n=5) | 79-91% | 87-96% | 95-103% | 85-91% |