| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.104 | 2.004 |
| 5.00 | 1.713 | 1.613 |
| 2.50 | 1.188 | 1.088 |
| 1.25 | 0.804 | 0.704 |
| 0.63 | 0.590 | 0.490 |
| 0.31 | 0.362 | 0.262 |
| 0.16 | 0.206 | 0.106 |
| 0.00 | 0.100 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-97% | 94% |
| EDTA plasma(n=5) | 86-113% | 100% |
| Heparin plasma(n=5) | 80-91% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-111% | 85-115% | 85-95% | 91-103% |
| EDTA plasma(n=5) | 91-98% | 89-114% | 93-114% | 86-101% |
| Heparin plasma(n=5) | 91-108% | 80-118% | 90-107% | 93-119% |