| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 150.00 | 2.248 | 2.166 |
| 75.00 | 1.675 | 1.593 |
| 37.50 | 1.174 | 1.092 |
| 18.75 | 0.892 | 0.810 |
| 9.38 | 0.589 | 0.507 |
| 4.69 | 0.365 | 0.283 |
| 2.35 | 0.234 | 0.152 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 94-107% | 100% |
| Heparin plasma(n=5) | 79-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-105% | 87-98% | 86-97% |
| EDTA plasma(n=5) | 82-93% | 81-96% | 79-91% | 83-94% |
| Heparin plasma(n=5) | 92-101% | 88-102% | 87-102% | 93-102% |