| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 2.204 | 2.132 |
| 2500.00 | 1.673 | 1.601 |
| 1250.00 | 1.162 | 1.090 |
| 625.00 | 0.813 | 0.741 |
| 312.50 | 0.525 | 0.453 |
| 156.25 | 0.336 | 0.264 |
| 78.13 | 0.234 | 0.162 |
| 0.00 | 0.072 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-97% | 89% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 80-92% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-95% | 95-105% | 88-102% | 93-101% |
| EDTA plasma(n=5) | 88-92% | 86-97% | 85-97% | 85-94% |
| Heparin plasma(n=5) | 85-100% | 84-98% | 95-101% | 80-96% |