| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.131 | 2.049 |
| 2000.00 | 1.527 | 1.445 |
| 1000.00 | 1.282 | 1.200 |
| 500.00 | 0.828 | 0.746 |
| 250.00 | 0.527 | 0.445 |
| 125.00 | 0.321 | 0.239 |
| 62.50 | 0.203 | 0.121 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 85-101% | 87-102% | 84-98% |
| EDTA plasma(n=5) | 83-96% | 93-102% | 95-103% | 93-102% |
| Heparin plasma(n=5) | 92-101% | 88-102% | 87-102% | 93-102% |