48 T | $265.00 |
96 T | $376.00 |
Concentration (U/mL) | OD | Corrected OD |
---|---|---|
50.00 | 2.096 | 2.003 |
25.00 | 1.527 | 1.434 |
12.50 | 1.225 | 1.132 |
6.25 | 0.861 | 0.768 |
3.13 | 0.577 | 0.484 |
1.57 | 0.318 | 0.225 |
0.79 | 0.236 | 0.143 |
0.00 | 0.093 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 84-93% | 89% |
Heparin plasma(n=5) | 97-107% | 102% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 90-96% | 87-1028% | 85-97% |
EDTA plasma(n=5) | 96-105% | 87-96% | 93-108% | 92-105% |
Heparin plasma(n=5) | 91-99% | 81-93% | 78-91% | 87-98% |