48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
8000.00 | 0.198 | |
4000.00 | 0.397 | |
2000.00 | 0.675 | |
1000.00 | 0.967 | |
500.00 | 1.168 | |
250.00 | 1.579 | |
125.00 | 1.863 | |
0.00 | 2.219 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 94-107% | 100% |
EDTA plasma(n=5) | 81-97% | 89% |
Heparin plasma(n=5) | 81-95% | 88% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 79-96% | 89-101% | 88-97% |
EDTA plasma(n=5) | 97-105% | 81-93% | 98-105% | 87-98% |
Heparin plasma(n=5) | 95-102% | 78-91% | 79-92% | 80-92% |