| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 25.00 | 2.045 | 1.959 |
| 12.50 | 1.682 | 1.596 |
| 6.25 | 1.205 | 1.119 |
| 3.13 | 0.921 | 0.835 |
| 1.57 | 0.556 | 0.470 |
| 0.79 | 0.324 | 0.238 |
| 0.40 | 0.236 | 0.150 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-94% | 87% |
| EDTA plasma(n=5) | 91-105% | 98% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 87-101% | 89-99% | 85-95% |
| EDTA plasma(n=5) | 96-101% | 85-92% | 78-92% | 92-103% |
| Heparin plasma(n=5) | 82-96% | 78-91% | 88-96% | 87-98% |