| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.258 | 2.171 |
| 5.00 | 1.663 | 1.576 |
| 2.50 | 1.109 | 1.022 |
| 1.25 | 0.871 | 0.784 |
| 0.63 | 0.557 | 0.470 |
| 0.32 | 0.336 | 0.249 |
| 0.16 | 0.256 | 0.169 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 79-96% | 85-99% | 79-93% |
| EDTA plasma(n=5) | 81-96% | 93-106% | 82-95% | 89-97% |
| Heparin plasma(n=5) | 95-105% | 85-92% | 96-103% | 87-98% |