| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.987 | 1.898 |
| 500.00 | 1.515 | 1.426 |
| 250.00 | 1.106 | 1.017 |
| 125.00 | 0.828 | 0.739 |
| 62.50 | 0.589 | 0.500 |
| 31.25 | 0.377 | 0.288 |
| 15.63 | 0.226 | 0.137 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-96% | 87% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-101% | 96-105% | 89-103% | 95-103% |
| EDTA plasma(n=5) | 93-102% | 89-97% | 85-94% | 93-102% |
| Heparin plasma(n=5) | 93-106% | 86-93% | 91-105% | 87-98% |