| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.154 | 2.065 |
| 5.00 | 1.631 | 1.542 |
| 2.50 | 1.043 | 0.954 |
| 1.25 | 0.921 | 0.832 |
| 0.63 | 0.614 | 0.525 |
| 0.32 | 0.357 | 0.268 |
| 0.16 | 0.226 | 0.137 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-98% | 90% |
| EDTA plasma(n=5) | 87-102% | 94% |
| Heparin plasma(n=5) | 95-107% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-101% | 89-97% | 85-94% | 87-98% |
| EDTA plasma(n=5) | 92-101% | 83-98% | 86-96% | 86-97% |
| Heparin plasma(n=5) | 88-95% | 92-101% | 96-105% | 93-101% |