| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.024 | 1.929 |
| 500.00 | 1.620 | 1.525 |
| 250.00 | 1.140 | 1.045 |
| 125.00 | 0.840 | 0.745 |
| 62.50 | 0.521 | 0.426 |
| 31.25 | 0.333 | 0.238 |
| 15.63 | 0.230 | 0.135 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-118% | 106% |
| EDTA plasma(n=5) | 86-97% | 92% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-116% | 89-106% | 81-89% | 84-102% |
| EDTA plasma(n=5) | 88-120% | 80-116% | 86-119% | 93-102% |
| Heparin plasma(n=5) | 95-104% | 85-100% | 94-120% | 83-97% |