| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.266 | 2.164 |
| 5.00 | 1.587 | 1.485 |
| 2.50 | 1.182 | 1.080 |
| 1.25 | 0.813 | 0.711 |
| 0.63 | 0.555 | 0.453 |
| 0.32 | 0.327 | 0.225 |
| 0.16 | 0.186 | 0.084 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-104% | 86-98% | 91-102% |
| EDTA plasma(n=5) | 79-91% | 87-96% | 92-105% | 98-107% |
| Heparin plasma(n=5) | 93-102% | 80-93% | 93-105% | 84-98% |