Products

Human sHLA-G(soluble human leucocyte antigen G) Microsample ELISA Kit
One-click to copy product information
ELK9364MS
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human sHLA-G(soluble human leucocyte antigen G) Microsample ELISA Kit
Reactivity: Human
Alternative Names: sHLA-G
Assay Type: Sandwich
Sensitivity: 0.066 ng/mL
Standard: 10 ng/mL
Detection Range: 0.16-10 ng/mL
Sample Type: serum, plasma, tissue homogenates and other biological fluids
Assay Length: 3.5h
Research Area: Apoptosis;Tumor immunity;Infection immunity;Immune molecule;Hematology;Reproductive science;Immunodeficiency;Autoimmunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human sHLA-G. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human sHLA-G. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human sHLA-G, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human sHLA-G in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
10.00 1.954 1.871
5.00 1.561 1.478
2.50 1.081 0.998
1.25 0.830 0.747
0.63 0.584 0.501
0.31 0.334 0.251
0.16 0.233 0.150
0.00 0.083 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant sHLA-G and the recovery rates were calculated by comparing the measured value to the expected amount of sHLA-G in samples.
Matrix Recovery range Average
serum(n=5) 81-100% 91%
EDTA plasma(n=5) 86-96% 91%
Heparin plasma(n=5) 88-109% 99%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sHLA-G and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 90-100% 91-108% 92-99% 91-108%
EDTA plasma(n=5) 83-101% 93-114% 92-117% 85-112%
Heparin plasma(n=5) 91-100% 94-117% 85-94% 82-106%

Product Components