| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 2.045 | 1.977 |
| 4000.00 | 1.628 | 1.560 |
| 2000.00 | 1.128 | 1.060 |
| 1000.00 | 0.827 | 0.759 |
| 500.00 | 0.487 | 0.419 |
| 250.00 | 0.369 | 0.301 |
| 125.00 | 0.258 | 0.190 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 96-108% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 82-93% | 82-96% | 92-103% |
| EDTA plasma(n=5) | 90-99% | 89-102% | 87-102% | 85-94% |
| Heparin plasma(n=5) | 87-101% | 95-102% | 89-98% | 95-103% |