| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.231 | 2.146 |
| 500.00 | 1.635 | 1.550 |
| 250.00 | 1.196 | 1.111 |
| 125.00 | 0.897 | 0.812 |
| 62.50 | 0.518 | 0.433 |
| 31.25 | 0.322 | 0.237 |
| 15.63 | 0.206 | 0.121 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-91% | 84% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 90-107% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-105% | 92-104% | 87-93% | 95-106% |
| EDTA plasma(n=5) | 88-103% | 81-98% | 82-96% | 87-105% |
| Heparin plasma(n=5) | 95-102% | 95-103% | 88-96% | 81-95% |