| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.126 | 2.040 |
| 5.00 | 1.673 | 1.587 |
| 2.50 | 1.194 | 1.108 |
| 1.25 | 0.869 | 0.783 |
| 0.63 | 0.573 | 0.487 |
| 0.32 | 0.365 | 0.279 |
| 0.16 | 0.189 | 0.103 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 84-102% | 93% |
| Heparin plasma(n=5) | 79-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-96% | 79-98% | 81-92% | 87-101% |
| EDTA plasma(n=5) | 89-99% | 85-95% | 93-102% | 93-106% |
| Heparin plasma(n=5) | 87-102% | 84-98% | 89-102% | 82-96% |