| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.255 | 2.170 |
| 5.00 | 1.527 | 1.442 |
| 2.50 | 1.175 | 1.090 |
| 1.25 | 0.828 | 0.743 |
| 0.63 | 0.567 | 0.482 |
| 0.32 | 0.389 | 0.304 |
| 0.16 | 0.175 | 0.090 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 80-96% | 88% |
| Heparin plasma(n=5) | 83-99% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 92-101% | 82-101% | 91-103% |
| EDTA plasma(n=5) | 87-98% | 87-102% | 84-98% | 89-102% |
| Heparin plasma(n=5) | 82-96% | 83-94% | 92-105% | 85-102% |