Products

Pig PIIINP(Procollagen III N-Terminal Propeptide) ELISA Kit
One-click to copy product information
ELK9263
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Pig PIIINP(Procollagen III N-Terminal Propeptide) ELISA Kit
Reactivity: Pig
Alternative Names: P3NP; N-Propeptide Of Type III Procollagen; Procollagen III Amino Terminal Propeptide
Assay Type: Sandwich
Sensitivity: 2.36 ng/mL
Standard: 400 ng/mL
Detection Range: 6.25-400 ng/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 3.5h
Research Area: Metabolic pathway;Hepatology;Bone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Pig PIIINP. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Pig PIIINP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Pig PIIINP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Pig PIIINP in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
400.00 2.035 1.948
200.00 1.506 1.419
100.00 1.140 1.053
50.00 0.933 0.846
25.00 0.518 0.431
12.50 0.313 0.226
6.25 0.161 0.074
0.00 0.087 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant PIIINP and the recovery rates were calculated by comparing the measured value to the expected amount of PIIINP in samples.
Matrix Recovery range Average
serum(n=5) 92-107% 99%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 81-97% 89%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PIIINP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 96-105% 95-103% 95-103% 84-98%
EDTA plasma(n=5) 93-102% 87-98% 87-98% 85-96%
Heparin plasma(n=5) 80-96% 84-98% 91-97% 81-93%