| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.096 | 2.008 |
| 250.00 | 1.668 | 1.580 |
| 125.00 | 1.223 | 1.135 |
| 62.50 | 0.933 | 0.845 |
| 31.25 | 0.577 | 0.489 |
| 15.63 | 0.318 | 0.230 |
| 7.82 | 0.236 | 0.148 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-107% | 102% |
| EDTA plasma(n=5) | 94-107% | 100% |
| Heparin plasma(n=5) | 82-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-96% | 85-98% | 95-104% | 93-102% |
| EDTA plasma(n=5) | 88-102% | 85-97% | 80-92% | 81-93% |
| Heparin plasma(n=5) | 87-101% | 96-105% | 89-103% | 95-103% |