| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.179 | 2.086 |
| 100.00 | 1.612 | 1.519 |
| 50.00 | 1.128 | 1.035 |
| 25.00 | 0.813 | 0.720 |
| 12.50 | 0.577 | 0.484 |
| 6.25 | 0.336 | 0.243 |
| 3.13 | 0.226 | 0.133 |
| 0.00 | 0.093 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-107% | 102% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-96% | 88-101% | 90-99% | 87-98% |
| EDTA plasma(n=5) | 95-102% | 86-97% | 88-95% | 87-98% |
| Heparin plasma(n=5) | 86-97% | 90-101% | 93-101% | 93-102% |