| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.131 | 2.044 |
| 1000.00 | 1.708 | 1.621 |
| 500.00 | 1.273 | 1.186 |
| 250.00 | 0.806 | 0.719 |
| 125.00 | 0.567 | 0.480 |
| 62.50 | 0.357 | 0.270 |
| 31.25 | 0.235 | 0.148 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 84-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-99% | 95-102% | 82-90% | 87-98% |
| EDTA plasma(n=5) | 93-101% | 91-99% | 83-97% | 87-98% |
| Heparin plasma(n=5) | 86-95% | 81-96% | 87-96% | 85-97% |