Product name: GSSG(Oxidized Glutathione) ELISA Kit
Reactivity: General
Alternative Names: Oxidized Glutathione
Assay Type: Competitive Inhibition
Sensitivity: 89.5 ng/mL
Standard: 20000 ng/mL
Detection Range: 312.5-20000 ng/mL
Sample Type: biological agents and other biological fluids
Assay Length: 2h
Research Area: Metabolic pathway
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with GSSG. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to GSSG. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of GSSG in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
20000.00 0.212
10000.00 0.352
5000.00 0.692
2500.00 0.898
1250.00 1.139
625.00 1.592
312.50 1.863
0.00 2.316

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant GSSG and the recovery rates were calculated by comparing the measured value to the expected amount of GSSG in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 80-96% 88%
Heparin plasma(n=5) 88-103% 95%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GSSG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-98% 89-102% 87-94% 85-96%
EDTA plasma(n=5) 93-102% 93-106% 87-101% 91-104%
Heparin plasma(n=5) 82-96% 95-102% 87-98% 81-97%