| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.058 | 1.957 |
| 100.00 | 1.556 | 1.455 |
| 50.00 | 1.196 | 1.095 |
| 25.00 | 0.847 | 0.746 |
| 12.50 | 0.594 | 0.493 |
| 6.25 | 0.387 | 0.286 |
| 3.13 | 0.218 | 0.117 |
| 0.00 | 0.101 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-107% | 98% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 80-96% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-105% | 88-96% | 87-98% | 93-101% |
| EDTA plasma(n=5) | 91-99% | 83-97% | 87-105% | 79-93% |
| Heparin plasma(n=5) | 94-106% | 83-97% | 93-102% | 89-97% |