| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.984 | 1.890 |
| 2500.00 | 1.507 | 1.413 |
| 1250.00 | 1.193 | 1.099 |
| 625.00 | 0.896 | 0.802 |
| 312.50 | 0.567 | 0.473 |
| 156.25 | 0.339 | 0.245 |
| 78.13 | 0.258 | 0.164 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-106% | 100% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-93% | 94-106% | 83-97% | 93-102% |
| EDTA plasma(n=5) | 89-97% | 85-97% | 87-96% | 83-94% |
| Heparin plasma(n=5) | 92-105% | 89-97% | 85-92% | 89-96% |