| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 8000.00 | 0.212 | |
| 4000.00 | 0.375 | |
| 2000.00 | 0.583 | |
| 1000.00 | 0.997 | |
| 500.00 | 1.139 | |
| 250.00 | 1.537 | |
| 125.00 | 1.755 | |
| 0.00 | 2.213 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 94-107% | 100% |
| EDTA plasma(n=5) | 84-98% | 91% |
| Heparin plasma(n=5) | 77-95% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-107% | 90-99% | 92-103% | 86-97% |
| EDTA plasma(n=5) | 83-96% | 81-94% | 86-98% | 92-101% |
| Heparin plasma(n=5) | 89-99% | 86-102% | 91-103% | 89-97% |