| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 300.00 | 0.199 | |
| 150.00 | 0.310 | |
| 75.00 | 0.555 | |
| 37.50 | 0.853 | |
| 18.75 | 1.122 | |
| 9.38 | 1.565 | |
| 4.69 | 1.877 | |
| 0.00 | 2.316 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 85-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-96% | 95-101% | 88-102% |
| EDTA plasma(n=5) | 79-92% | 85-96% | 80-98% | 95-102% |
| Heparin plasma(n=5) | 93-101% | 86-92% | 81-93% | 89-105% |