48 T | $265.00 |
96 T | $376.00 |
Concentration (uIU/mL) | OD | Corrected OD |
---|---|---|
12.00 | 2.043 | 1.960 |
6.00 | 1.587 | 1.504 |
3.00 | 1.227 | 1.144 |
1.50 | 0.958 | 0.875 |
0.75 | 0.532 | 0.449 |
0.38 | 0.357 | 0.274 |
0.19 | 0.161 | 0.078 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-95% | 90% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 84-102% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-92% | 97-103% | 87-98% | 95-102% |
EDTA plasma(n=5) | 87-102% | 82-91% | 86-93% | 81-94% |
Heparin plasma(n=5) | 87-97% | 86-92% | 89-103% | 89-98% |