| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.027 | 1.954 |
| 1000.00 | 1.720 | 1.647 |
| 500.00 | 1.227 | 1.154 |
| 250.00 | 0.907 | 0.834 |
| 125.00 | 0.587 | 0.514 |
| 62.50 | 0.324 | 0.251 |
| 31.25 | 0.135 | 0.062 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-95% | 88% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |
| EDTA plasma(n=5) | 95-102% | 87-96% | 78-92% | 85-97% |
| Heparin plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |