| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.266 | 2.180 |
| 500.00 | 1.638 | 1.552 |
| 250.00 | 1.052 | 0.966 |
| 125.00 | 0.897 | 0.811 |
| 62.50 | 0.509 | 0.423 |
| 31.25 | 0.335 | 0.249 |
| 15.63 | 0.205 | 0.119 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 90-105% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-91% | 89-103% | 82-95% | 86-99% |
| EDTA plasma(n=5) | 87-95% | 82-94% | 95-102% | 87-103% |
| Heparin plasma(n=5) | 81-93% | 89-104% | 78-96% | 79-98% |