| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.211 | |
| 500.00 | 0.467 | |
| 250.00 | 0.696 | |
| 125.00 | 0.967 | |
| 62.50 | 1.119 | |
| 31.25 | 1.522 | |
| 15.63 | 1.871 | |
| 0.00 | 2.132 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-90% | 84% |
| EDTA plasma(n=5) | 84-97% | 91% |
| Heparin plasma(n=5) | 89-105% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-94% | 85-103% | 89-97% | 79-95% |
| EDTA plasma(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
| Heparin plasma(n=5) | 85-90% | 84-99% | 85-97% | 88-102% |