| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.270 | 2.187 |
| 5.00 | 1.536 | 1.453 |
| 2.50 | 1.125 | 1.042 |
| 1.25 | 0.836 | 0.753 |
| 0.63 | 0.539 | 0.456 |
| 0.32 | 0.315 | 0.232 |
| 0.16 | 0.174 | 0.091 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 80-97% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 83-94% | 85-98% | 85-97% |
| EDTA plasma(n=5) | 91-102% | 86-102% | 87-99% | 99-105% |
| Heparin plasma(n=5) | 87-102% | 82-91% | 86-93% | 81-94% |