| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.961 | 1.867 |
| 5.00 | 1.628 | 1.534 |
| 2.50 | 1.027 | 0.933 |
| 1.25 | 0.896 | 0.802 |
| 0.63 | 0.583 | 0.489 |
| 0.32 | 0.366 | 0.272 |
| 0.16 | 0.214 | 0.120 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 92% |
| EDTA plasma(n=5) | 88-107% | 94% |
| Heparin plasma(n=5) | 97-105% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
| EDTA plasma(n=5) | 92-101% | 82-98% | 89-97% | 85-93% |
| Heparin plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |