| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.941 | 1.860 |
| 1000.00 | 1.412 | 1.331 |
| 500.00 | 1.036 | 0.955 |
| 250.00 | 0.861 | 0.780 |
| 125.00 | 0.496 | 0.415 |
| 62.50 | 0.312 | 0.231 |
| 31.25 | 0.175 | 0.094 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 93-107% | 100% |
| EDTA plasma(n=5) | 85-99% | 92% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-98% | 92-104% | 82-96% | 95-102% |
| EDTA plasma(n=5) | 85-97% | 87-96% | 85-96% | 85-95% |
| Heparin plasma(n=5) | 86-100% | 93-108% | 80-101% | 87-96% |