| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.038 | 1.950 |
| 1000.00 | 1.586 | 1.498 |
| 500.00 | 1.106 | 1.018 |
| 250.00 | 0.871 | 0.783 |
| 125.00 | 0.509 | 0.421 |
| 62.50 | 0.342 | 0.254 |
| 31.25 | 0.230 | 0.142 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 94-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
| EDTA plasma(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |
| Heparin plasma(n=5) | 86-96% | 93-104% | 89-97% | 85-94% |