| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (mIU/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 2.201 | 2.116 |
| 25.00 | 1.515 | 1.430 |
| 12.50 | 1.194 | 1.109 |
| 6.25 | 0.933 | 0.848 |
| 3.13 | 0.563 | 0.478 |
| 1.57 | 0.297 | 0.212 |
| 0.79 | 0.319 | 0.234 |
| 0.00 | 0.085 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-92% | 85% |
| EDTA plasma(n=5) | 92-107% | 99% |
| Heparin plasma(n=5) | 87-99% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-107% | 83-99% | 80-95% | 87-96% |
| EDTA plasma(n=5) | 85-95% | 79-96% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 95-104% | 87-96% | 92-101% | 82-98% |