Products

Tyr(Tyrosol) ELISA Kit
One-click to copy product information
ELK8270
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: Tyr(Tyrosol) ELISA Kit
Reactivity: General
Alternative Names: Tyrosol
Assay Type: Competitive Inhibition
Sensitivity: 136.2 ng/mL
Standard: 30000 ng/mL
Detection Range: 468.75-30000 ng/mL
Sample type: biological agents and other biological fluids
Assay length: 2h
Research Area: Enzyme & Kinase;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Tyr. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Tyr. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Tyr in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
30000.00 0.191
15000.00 0.357
7500.00 0.577
3750.00 0.842
1875.00 1.275
937.50 1.578
468.75 1.769
0.00 2.265

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Tyr and the recovery rates were calculated by comparing the measured value to the expected amount of Tyr in samples.
Matrix Recovery range Average
serum(n=5) 90-105% 97%
EDTA plasma(n=5) 85-97% 91%
Heparin plasma(n=5) 81-95% 99%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Tyr and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 96-103% 89-97% 91-103%
EDTA plasma(n=5) 85-94% 79-96% 92-101% 82-90%
Heparin plasma(n=5) 85-94% 87-96% 78-93% 82-90%