| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 600.00 | 0.212 | |
| 300.00 | 0.375 | |
| 150.00 | 0.611 | |
| 75.00 | 0.897 | |
| 37.50 | 1.197 | |
| 18.75 | 1.511 | |
| 9.38 | 1.813 | |
| 0.00 | 2.325 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-93% | 86% |
| Heparin plasma(n=5) | 93-107% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-106% | 86-99% | 79-93% | 89-97% |
| EDTA plasma(n=5) | 87-98% | 85-96% | 92-102% | 81-93% |
| Heparin plasma(n=5) | 89-102% | 87-103% | 97-104% | 82-97% |