| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 2.214 | 2.125 |
| 20.00 | 1.387 | 1.298 |
| 10.00 | 1.073 | 0.984 |
| 5.00 | 0.892 | 0.803 |
| 2.50 | 0.539 | 0.450 |
| 1.25 | 0.379 | 0.290 |
| 0.63 | 0.202 | 0.113 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-95% | 89% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 86-92% | 81-93% | 89-104% |
| EDTA plasma(n=5) | 96-101% | 85-92% | 79-96% | 89-101% |
| Heparin plasma(n=5) | 87-98% | 92-104% | 82-96% | 95-102% |