| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.950 | 1.855 |
| 5.00 | 1.579 | 1.484 |
| 2.50 | 0.913 | 0.818 |
| 1.25 | 0.726 | 0.631 |
| 0.63 | 0.576 | 0.481 |
| 0.32 | 0.369 | 0.274 |
| 0.16 | 0.216 | 0.121 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-102% | 96-105% | 85-93% | 89-97% |
| EDTA plasma(n=5) | 89-97% | 85-97% | 87-96% | 85-96% |
| Heparin plasma(n=5) | 95-104% | 93-102% | 79-94% | 87-98% |