| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.154 | 2.067 |
| 500.00 | 1.556 | 1.469 |
| 250.00 | 1.135 | 1.048 |
| 125.00 | 0.861 | 0.774 |
| 62.50 | 0.518 | 0.431 |
| 31.25 | 0.365 | 0.278 |
| 15.63 | 0.187 | 0.100 |
| 0.00 | 0.087 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 83-95% | 89% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 82-95% | 92-105% |
| EDTA plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |
| Heparin plasma(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |