Products

DHVD3(1,25-Dihydroxyvitamin D3) ELISA Kit
One-click to copy product information
ELK8091
Size: Price:
48 T $265.00
96 T $376.00

Overview

Product name: DHVD3(1,25-Dihydroxyvitamin D3) ELISA Kit
Reactivity: General
Alternative Names: DH-VD3; Calcitriol; Rocaltrol; Calcijex; 1,25-(OH)2D3; 1,25-Dihydroxycholecalciferol; 1,25-Dihydroxy Vitamin D3
Assay Type: Competitive Inhibition
Sensitivity: 48.8 pg/mL
Standard: 10000 pg/mL
Detection Range: 156.25-10000 pg/mL
Sample type: serum, plasma and other biological fluids
Assay length: 2h
Research Area: Metabolic pathway;Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with DHVD3. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to DHVD3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of DHVD3 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
10000.00 0.189
5000.00 0.386
2500.00 0.596
1250.00 0.878
625.00 1.169
312.50 1.592
156.25 1.796
0.00 2.112

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant DHVD3 and the recovery rates were calculated by comparing the measured value to the expected amount of DHVD3 in samples.
Matrix Recovery range Average
serum(n=5) 80-97% 88%
EDTA plasma(n=5) 82-98% 90%
Heparin plasma(n=5) 82-94% 90%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of DHVD3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 82-94% 85-96% 82-96% 95-102%
EDTA plasma(n=5) 87-98% 92-105% 83-92% 89-98%
Heparin plasma(n=5) 81-94% 81-96% 93-106% 82-95%