| 48 T | $412.00 | 
| 96 T | $588.00 | 
| 96 T*5 | $2499.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 200.00 | 2.204 | 2.120 | 
| 100.00 | 1.542 | 1.458 | 
| 50.00 | 1.176 | 1.092 | 
| 25.00 | 0.828 | 0.744 | 
| 12.50 | 0.457 | 0.373 | 
| 6.25 | 0.357 | 0.273 | 
| 3.13 | 0.175 | 0.091 | 
| 0.00 | 0.084 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 84-103% | 96% | 
| EDTA plasma(n=5) | 82-96% | 87% | 
| Heparin plasma(n=5) | 86-97% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-93% | 95-104% | 82-96% | 85-98% | 
| EDTA plasma(n=5) | 93-102% | 78-92% | 87-98% | 80-93% | 
| Heparin plasma(n=5) | 88-97% | 86-92% | 95-103% | 98-106% |