| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.126 | 2.054 |
| 5.00 | 1.583 | 1.511 |
| 2.50 | 1.198 | 1.126 |
| 1.25 | 0.856 | 0.784 |
| 0.63 | 0.547 | 0.475 |
| 0.32 | 0.302 | 0.230 |
| 0.16 | 0.236 | 0.164 |
| 0.00 | 0.072 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 93-107% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |
| EDTA plasma(n=5) | 85-102% | 81-96% | 87-105% | 83-92% |
| Heparin plasma(n=5) | 88-97% | 86-92% | 79-103% | 89-97% |