| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 600.00 | 2.154 | 2.086 |
| 300.00 | 1.674 | 1.606 |
| 150.00 | 1.174 | 1.106 |
| 75.00 | 0.852 | 0.784 |
| 37.50 | 0.518 | 0.450 |
| 18.75 | 0.262 | 0.194 |
| 9.38 | 0.205 | 0.137 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 92-107% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-96% | 85-97% | 89-101% | 85-97% |
| EDTA plasma(n=5) | 86-93% | 91-96% | 87-98% | 95-102% |
| Heparin plasma(n=5) | 84-99% | 83-95% | 83-95% | 96-101% |