| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.268 | 2.191 |
| 1000.00 | 1.612 | 1.535 |
| 500.00 | 1.142 | 1.065 |
| 250.00 | 0.806 | 0.729 |
| 125.00 | 0.453 | 0.376 |
| 62.50 | 0.315 | 0.238 |
| 31.25 | 0.256 | 0.179 |
| 0.00 | 0.077 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-103% | 96% |
| EDTA plasma(n=5) | 78-93% | 85% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-97% | 87-96% | 86-98% | 91-103% |
| EDTA plasma(n=5) | 83-94% | 80-92% | 86-102% | 87-103% |
| Heparin plasma(n=5) | 81-97% | 88-102% | 85-93% | 83-97% |