| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.154 | 2.068 |
| 5.00 | 1.527 | 1.441 |
| 2.50 | 1.165 | 1.079 |
| 1.25 | 0.895 | 0.809 |
| 0.63 | 0.517 | 0.431 |
| 0.32 | 0.273 | 0.187 |
| 0.16 | 0.189 | 0.103 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-97% | 89% |
| EDTA plasma(n=5) | 84-99% | 91% |
| Heparin plasma(n=5) | 79-91% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 83-96% | 97-105% | 85-97% |
| EDTA plasma(n=5) | 88-102% | 81-96% | 86-93% | 82-95% |
| Heparin plasma(n=5) | 88-105% | 79-93% | 87-98% | 85-96% |