| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.925 | 1.842 |
| 1000.00 | 1.562 | 1.479 |
| 500.00 | 1.073 | 0.990 |
| 250.00 | 0.892 | 0.809 |
| 125.00 | 0.543 | 0.460 |
| 62.50 | 0.312 | 0.229 |
| 31.25 | 0.165 | 0.082 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 96-107% | 101% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 85-97% | 91% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 97-105% | 92-101% | 85-92% |
| EDTA plasma(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
| Heparin plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |