| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.401 | 2.328 |
| 1000.00 | 1.563 | 1.490 |
| 500.00 | 1.128 | 1.055 |
| 250.00 | 0.897 | 0.824 |
| 125.00 | 0.576 | 0.503 |
| 62.50 | 0.313 | 0.240 |
| 31.25 | 0.196 | 0.123 |
| 0.00 | 0.073 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 97-107% | 102% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-102% | 93-101% | 97-105% | 95-103% |
| EDTA plasma(n=5) | 86-93% | 91-101% | 87-98% | 95-104% |
| Heparin plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |